Technical Information
RS 232/RS 485 Communication Description
17
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
193/224
NOTE
Using the RS 485 interface, the message structure re-mains identical to the one
used for the RS 232 interface, the only difference being that the value assigned
to the ADDRESS <ADDR>
The controller can answers with the following response types:
Table 5
Response types
Type Length
Value
Description
Logic
1 byte
-
After a read instruction of a logic
window
Numeric
6 bytes
-
After a read instruction of a numeric
window
Alphanumeric
10 bytes
-
After a read instruction of an
alphanumeric window
ACK
1 byte
(0x6)
The command execu-tion has been
successfully completed
NACK
1 byte
(0x15)
The command execu-tion has been
failed
Unknown
Window
1 byte
(0x32)
The specified window in the
command is not a valid window
Data Type Error 1 byte
(0x33)
The data type specified in the
command (Logic, Numeric or
Alphanu-meric) is not accorded with
the specified Win-dow
Out of Range
1 byte
0x34)
The value expressed during a write
com-mand is out of the range value
of the specified window
Win Disabled
1 byte
(0x35)
The specified window is Read Only
or tempo-rarily disabled (for ex-
ample you can’t write the Soft Start
when the Pump is running)