TriScroll 620 Series Vacuum Pump
DRAFT 10/30/14
Troubleshooting
Table 5 contains a list of possible problems, their probable causes, and corrective actions.
Table 5 Troubleshooting Chart
Problem
Probable Cause
Corrective Action
Pump won’t start
Circuit breaker open
Close breaker.
Identify cause of overload.
Motor thermal protector open
Allow motor to cool.
Identify cause of overload.
Electrical short under the motor
electrical cover
Inspect and repair.
Wiring loose or cut
Repair or replace.
Excessive voltage drop
Check size and length of power supply
cable.
Defective motor
Inspect. Contact Agilent.
Poor ultimate pressure
System leak
Locate and repair leak.
Water in pump
Flush pump with air or dry nitrogen.
Gas ballast plugged
Replace breather vent. Contact Agilent.
Solvent in pump
Flush pump with air or dry nitrogen.
Install trap or filter.
Seals worn out
Replace tip seals. (Table 7 and Table 8
on page 18 list maintenance kits and
service options.)
Poor conductance to pump
Replumb with shorter and/or larger
diameter tubing.
Pump makes hammering noise
Pump overheated
Check ambient temperature.
Check ventilation to pump.
Debris in pump
Check inlet screen.
Flush pump.
Disassemble pump and inspect.
(Table 7 and Table 8 on page 18 list
maintenance kits and service options.)
Summary of Contents for TriScroll 620 Series
Page 4: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 6: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 20: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank DRAFT 10 30 14...
Page 24: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...
Page 25: ......