![Agilent Technologies SPS 4 User Manual Download Page 46](http://html1.mh-extra.com/html/agilent-technologies/sps-4/sps-4_user-manual_2867793046.webp)
46
Agilent SPS 4 Autosampler User’s Guide
3
Installation
Two Drain Arrangements
The waste rinse solution drains from the wash reservoir by means of a
gravity drain, which is the standard arrangement for draining the wash
reservoir. If preferred, you can create a pumped drain arrangement. This
section introduces the gravity drain arrangement and the pumped drain
arrangement.
Gravity Drain Arrangement
In a gravity drain arrangement, the rinse solution moves through a
channel (and two channels for dual port wash reservoir) of the peristaltic
pump to the inlet of the wash reservoir, as shown in
. It then
drains out from the outlet of the wash reservoir into the waste container
by means of a gravity drain.
Figure 16
Gravity Drain Arrangement
Pumped Drain Arrangement
In a pumped drain arrangement, the rinse solution moves through a
channel (or two channels for dual port wash reservoir) of the peristaltic
pump to the inlet of the wash reservoir. It then drains out from the outlet
of the wash reservoir into the waste container through other channel of
Dual Port
Wash Reservoir
OUT
IN
OUT
IN
IN
OUT
Wash Pump
Flow
Rinse Solution In
Waste Out
OUT
IN
IN
OUT
Flow
Rinse Solution In
Waste Out
Single Port
Wash Reservoir
Wash Pump
(B) For Dual Port Wash Reservoir
(A) For Single Port Wash Reservoir
FWD
ROTATION
FWD
ROTATION
FWD
ROTATION
Summary of Contents for SPS 4
Page 1: ...Agilent Technologies Agilent SPS 4 Autosampler User s Guide...
Page 28: ...28 Agilent SPS 4 Autosampler User s Guide 2 Introduction This page is intentionally left blank...
Page 52: ...52 Agilent SPS 4 Autosampler User s Guide 3 Installation This page is intentionally left blank...
Page 62: ...62 Agilent SPS 4 Autosampler User s Guide 5 Maintenance This page is intentionally left blank...
Page 96: ...96 Agilent SPS 4 Autosampler User s Guide A Appendices This page is intentionally left blank...
Page 97: ......