Installation
3
Agilent SPS 4 Autosampler User’s Guide
43
6
Slip the tip of the sample probe into the hole in the middle of the
Z- axis slide, through the mount nut and the accessories plate until the
probe mount block is sitting in the z- axis slide.
7
Slide the mount nut up onto the probe mount block and tighten to
secure the sample probe to the Z- axis slide.
8
Route the sample line from the sample probe to the left around the
probe arm housing. When using a carbon fiber sample probe, do not
run the sample line through the hook (or tube restraint) at the top of
the Z- axis slide. Connect the cable tie to the eyelet on the underside of
the probe arm cover and leave a loop of approximately 10 mm. Thread
the sample line through the loop in the cable tie as shown in
.
Figure 14
Routing the sample line tubing
N O T E
Do not use the tube restraint for pre- formed sample probes. It is only
used for non pre- formed sample probes.
Cable tie tubing through
the eyelet on bottom of
housing
Tube Anchor
Grommet
Colored Markers
Summary of Contents for SPS 4
Page 1: ...Agilent Technologies Agilent SPS 4 Autosampler User s Guide...
Page 28: ...28 Agilent SPS 4 Autosampler User s Guide 2 Introduction This page is intentionally left blank...
Page 52: ...52 Agilent SPS 4 Autosampler User s Guide 3 Installation This page is intentionally left blank...
Page 62: ...62 Agilent SPS 4 Autosampler User s Guide 5 Maintenance This page is intentionally left blank...
Page 96: ...96 Agilent SPS 4 Autosampler User s Guide A Appendices This page is intentionally left blank...
Page 97: ......