33
Memory Clearing, Sanitization and Removal Procedures
Force Internal to SD Card
NOTE
The Erase and Sanitize All operation resets the instrument’s LAN settings to their
factory defaults. For more details, see
“LAN Setup” on page 34
.
Force Internal to SD Card
This function is available only in the MXG Series B and EXG instruments.
The options for this setting are
Off
or
On
:
•
On
: Forces all internal settings and files to be stored only on the external SD memory card, including
instrument states, user data files, sweep list files, and other user created files. If it becomes necessary to
sanitize the instrument, these settings may then be securely retained by removing the SD card and
retaining it in the secure area.
•
Off
: All internal settings are stored in the instrument’s Main Memory (see
“MXG (Series B) & EXG
Memory” on page 16
). This memory may be sanitized using the
Erase and Sanitize All
function.
Model
Key Sequence
MXG (Series A/B)
EXG
File
>
More
>
Security
>
Erase and Sanitize All
>
Confirm Sanitize
PSG
ESG
Utility
>
Memory Catalog
>
More
>
Security
>
Erase and Sanitize All
>
Confirm Sanitize
Model
Key Sequence
MXG (Series B)
EXG
File
>
More
>
Security
>
Force Internal to SD Card
Summary of Contents for N5171B
Page 4: ...4 ...
Page 6: ...6 Contents ...
Page 8: ...8 Contacting Agilent Sales and Service Offices ...
Page 12: ...12 Products Covered by this Document Recommended Software Upgrades ...
Page 14: ...14 Security Terms and Definitions ...
Page 38: ...38 Using Secure Display ...
Page 42: ...42 Security Issues for Certain Firmware Revisions Recovering Erased System Files ...
Page 46: ...46 Procedure for Declassifying a Faulty Instrument ...