31
N2793A Performance Verification Procedures
The following procedure can be used to test the N2793A high- voltage
differential probe’s DC differential gain accuracy and bandwidth, which are
warranted specifications.
Table 10 Required Test Equipment
Description
Critical Specifications
Recommended Model
Part Number
Functions
Digitizing Oscilloscope
Bandwidth: >1 GHz
1 M
Ω
/50
Ω
selectable input
Agilent MSO9254A
Display probe output
Signal Generator
Precision DC voltage source
Amplitude accuracy: less
than or equal to 0.25%
1 M
Ω
/50
Ω
selectable load
Sine wave greater than or
equal to 800 MHz
Fluke 9500B Or
Agilent 1EU
option
Signal source for DC gain
and bandwidth
Performance Check Fixture
50
Ω
impedance controlled
microstrip trace
Agilent E2655B
Probing trace for
bandwidth test
Terminator
50
Ω
matched load
Termination for the PV
fixture
BNC Adapter
BNC (f) to SMA (m)
Interconnection between
calibrator and PV fixture
BNC Adapter
BNC(f) to Dual Banana (m)
adapter
Agilent 1251-2277
Termination between
probe and calibrator