28
Procedure
1
Configure the oscilloscope to the following settings:
Amplitude Scale (Channel 1): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 mV/div
Amplitude Scale (Channel 2): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mV/div
Time Scale: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200
μ
s/div
Acquisition mode: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 averages
Input impedance: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Ω
Trigger: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Channel 2
2
On the 9500B calibrator, configure
CH2
to the following settings:
Type: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TRIGGER
Load: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Ω
3
On the 9500B, connect the
CH2
active head to channel 2 on the
oscilloscope. Connect the
CH1
active head to channel 1 on the oscilloscope
4
On the N9500B, set
CH1
to the following settings and enable the output:
Waveform: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . square wave
Load: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MΩ
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1V pk-pk
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
5
On the oscilloscope, select the amplitude measurement and record the DC
amplitude (approximately 100 mV) of the square wave. This measurement is
only the oscilloscope.
6
Disable the 9500B’s output and disconnect the active head from channel 1
of the oscilloscope.
7
Connect the N2792A probe output to channel 1 of the oscilloscope.
8
Attach the BNC adapter to the 9500B’s
CH1
active head.
9
Attach the differential probe input leads by clipping the alligator clamp to
the BNC adapter banana post.
10
On the N9500B, configure
CH1
to to the following settings and enable the
output:
Waveform: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . square wave
Load: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MΩ
Amplitude: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1V
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz