Using Functions
3
N9310A User’s Guide
45
Generating a Pulse Modulated Signal
The signal generator generates pulse modulated signals with
the following characters:
Operation Example
Assume that you are going to generate an pulse modulated
signal with the following characters:
• Carrier frequency of 1 GHz, amplitude of –10 dBm
• Pulse period of 10 ms
• Pulse width of 6 ms
• Internal pulse source (default by
)
Refer to the following to configure your signal generator:
1
Pressing
presets the signal generator
2
Pressing
> 1 > GHz
sets the CW frequency to 1 GHz
3
Pressing
> –10 >
dBm
sets
the
CW amplitude to –10 dBm
4
Pressing
enters the Pulse modulation submenu
5
Pressing
Pulse Period > 10 > ms
sets pulse period to 10 ms
6
Pressing
Pulse width > 6 > ms
sets pulse width to 6 ms
7
Pressing
Pulse On/Off
enables pulse modulation
8
Pressing
On
enables pulse modulated signal output
For key reference, please refer to
Character
Range
Default
Pulse Source
Internal source (INT)
External source (EXT)
INT
Pulse Period
200
μ
s to 2 s (INT and EXT) 200
m
s, with 1
m
s resolution
Pulse Width
100
μ
s to 1 s (INT and EXT) 100
m
s, with 1
m
s resolution
Preset
Preset
Frequency
Amplitude
Pulse
RF
On/Off
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...