Agilent InfinityLab LC Series Diode Array Detectors User Manual
31
2
Site Requirements and Specifications
Performance Specifications
Flow cells
User-exchangeable, self-aligning cartridge cells with RFID tags.
Max-Light Cartridge Cell (Standard): 10 mm,
V
= 1.0 µL
Max-Light Cartridge Cell (High Sensitivity): 60 mm,
V
= 4 µL
Max-Light Cartridge Ultra Low Dispersion (ULD) Cell: 10 mm,
V
= 0.6 µL
Max-Light Cartridge High Dynamic Range (HDR) Cell: 3.7 mm,
V
= 0.9 µL
Max-Light Cartridge Cell LSS: 10 mm,
V
= 1.0 µL
Maximum Operating Pressure (MOP)
1
: 70 bar (60 bar for LSS)
Maximum Incidental Pressure (MIP)
2
: 150 bar
Spectral tools
Data analysis software for spectra evaluation, including spectral libraries
and peak purity functions
Analog output
Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output
Instrument control
Lab Advisor B.02.06 or above
LC and CE Drivers A.02.11 or above
For details about supported software versions refer to the compatibility
matrix of your version of the LC and CE Drivers
Local control
Agilent Instant Pilot (G4208A) B.02.19 or above
Communication
LAN, Controller Area Network (CAN), USB
ERI: ready, start, stop and shut-down signals
GLP
RFID for electronics records of flow cell and UV lamp conditions (path
length, volume, product number, serial number, test passed, usage)
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of lamp burn time with user settable limits and feedback
messages. Electronic records of maintenance and errors. Verification of
wavelength accuracy with deuterium lines.
Safety and mainte
-
nance
Extensive diagnostics, error detection and display through Agilent Instant
Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in major
maintenance areas.
Housing
All materials recyclable.
Others
Second generation of Electronic temperature control (ETC) for the complete
optical unit.
1
Maximum operating pressure (MOP): Maximum pressure at which a system can operate continuously un
-
der normal conditions.
2
Maximum incidental pressure (MIP): The maximum pressure which the system can experience during a
short time.
Table 2
Performance Specifications G7117A
Type
Specification