Agilent InfinityLab LC Series Diode Array Detectors User Manual
13
1
Introduction
Product Description
Product Description G7117C
The Agilent 1260 Infinity II Diode Array Detector HS with fixed slit is based on the
Agilent Max-Light cartridge cell with optofluidic waveguides that improve light
transmission to near 100% efficiency without sacrificing resolution caused by
cell dispersions effects.
With typical detector noise levels of <
±
0.6 µAU/cm the 60 mm flow cell gives up
to 10 times higher sensitivity than detectors with conventional flow cells.
Figure 3
Overview of the detector
Any compromising refractive index and thermal effects are almost completely
eliminated, resulting in significantly less baseline drift for more reliable and
precise peak integration.
For fast separations, this detector has multiple wavelength and full spectral
detection at sampling rates up to 120 Hz.
Status indicator
Lamp housing cover
Cartridge Flow Cell
Power switch
Leak drain
Air inlet