Agilent InfinityLab LC Series Diode Array Detectors User Manual
119
7
Error Information
Detector Error Messages
Wavelength Recalibration Lost
Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If no
data is available in the spectrometer ROM, the error message is generated.
Illegal Temperature Value from Sensor on Main
Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a value
outside the allowed range. The parameter of this event equals the measured
temperature in 1/100 centigrade. As a result the temperature control is switched
off.
Probable cause
Suggested actions
1
The detector is new.
Recalibrate the detector.
2
The detector has been repaired.
Please contact your Agilent service representa
-
tive.
Probable cause
Suggested actions
1
Defective sensor or main board.
Please contact your Agilent service representa
-
tive.
2
Detector is exposed to illegal ambient condi
-
tions.
Verify that the ambient conditions are within the
allowed range.