96
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
To Switch from the Standard or Inert EI ion source to the CI Ion Source
Procedure
1
Vent the MSD. See
2
Open the analyzer. See
“To Open the Analyzer Chamber”
3
Remove the EI ion source. See
4
Install the CI ion source. See
“To Install the CI Ion Source”
5
Install the CI/Extractor interface tip seal (p/n G3870-20542). See
Install the CI/Extractor Interface Tip Seal”
6
Close the analyzer. See
“To Close the Analyzer Chamber”
7
Pump down the MSD. See
“To Pump Down the MSD in CI Mode”
C A U T I O N
Always verify MSD performance in EI before switching to CI operation.
Always set up the CI MSD in PCI first, even if you are going to run NCI.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap.
See
Take
antistatic
precautions
before
you open
the analyzer chamber.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......