5977 Series MSD Operation Manual
3
Operating in Electron Ionization (EI) Mode
To Open the Analyzer Chamber
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Wrist strap, antistatic
•
Small (9300-0969)
•
Medium (9300-1257)
•
Large (9300-0970)
Procedure
1
Vent the MSD (see
2
Disconnect the side board control cable and the source power cable from
the side board.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap
and take other antistatic precautions (
) before opening the analyzer chamber.
Thumbscrews
Side board cable
Source power cable