General Maintenance
5
5977 Series MSD Operation Manual
123
Chemical residue
Only a small portion of your sample is ionized by the ion source. The majority
of any sample passes through the ion source without being ionized. It is
pumped away by the vacuum system. As a result, the exhaust from the foreline
pump will contain traces of the carrier gas and your samples. Exhaust from
the standard foreline pump also contains tiny droplets of foreline pump oil.
An oil trap is supplied with the standard foreline pump. This trap stops
only
pump oil droplets. It
does not
trap any other chemicals. If you are using toxic
solvents or analyzing toxic chemicals, do not use this oil trap. For all foreline
pumps, install a hose to take the exhaust from the foreline pump outdoors or
into a fume hood vented to the outdoors. For the standard foreline pump, this
requires removing the oil trap. Be sure to comply with your local air quality
regulations.
The fluids in the diffusion pump and standard foreline pump also collect
traces of the samples being analyzed. All used pump fluid should be
considered hazardous and handled accordingly. Dispose of used fluid
correctly, as specified by your local regulations.
Electrostatic discharge
All of the printed circuit boards in the MSD contain components that can be
damaged by electrostatic discharge (ESD). Do not handle or touch these
boards unless absolutely necessary. In addition, wires, contacts, and cables
can conduct ESD to the electronics boards to which they are connected. This is
especially true of the mass filter (quadrupole) contact wires which can carry
WA R N I N G
The oil trap supplied with the standard foreline pump stops only foreline pump oil. It
does not trap or filter out toxic chemicals. If you are using toxic solvents or
analyzing toxic chemicals, remove the oil trap. Do not use the trap if you have a CI
MSD. Install a hose to take the foreline pump exhaust outside or to a fume hood.
WA R N I N G
When replacing pump fluid, use appropriate chemical-resistant gloves and safety
glasses. Avoid all contact with the fluid.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......