CI Maintenance
6
5977 Series MSD Operation Manual
155
To Install the CI/Extractor Interface Tip Seal
Materials needed
•
Interface tip seal (G1999-60412)
The interface tip seal must be in place for the CI ion source and the
Extraction source.
Procedure
1
Verify that the EI extractor ion source or CI ion source is installed. This tip
seal must not be installed when a standard EI SST source or EI Inert
source is installed.
2
Remove the CI/Extractor tip seal from the ion source storage box and place
it over the end of the interface.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap
and take other antistatic precautions
before
you open the analyzer chamber.
CI/Extractor tip seal