Installation
Assembling the Foreline Hose
16
Agilent G6697A Oil Free IDP-10 Dry Scroll Pump Installation Guide
Assembling the Foreline Hose
To assemble the foreline hose:
1
Attach the KF25 Flange × 5/8 in. ID Hose barb adapter (G1099-20532) to one
end of the new foreline hose (G7077-60119).
2
Secure the barb fitting with 1x Clamp, Hose .437 in. DIA .56 in. Hose Clamp
(1400-3241) as shown in
a
Spray the KF25 hose adapter (p/n G1099-20532) with isopropyl alcohol
(IPA) solvent.
b
Place the KF25 hose adapter connector face on a cloth on your work
surface.
c
Slide the end of the foreline hose over the KF25 hose adapter.
d
Secure the KF25 hose adapter in the foreline hose by positioning the
hose clamp over the inserted portion of the adapter, and then using a
hand tool, tighten hand tight.
CAUTION
Do Not use any existing vacuum hoses or fittings. Any retained oil or
contamination from previously used hoses or fittings will damage the pump,
and may result in permanent pump failure.
Figure 5.
Attach barb fitting to new foreline hose and secure with hose clamp
CAUTION
Be careful not to scratch the face of the hose adapter. Deep grooves or
scratches may prevent a leak-free installation.