Control Panel and Indicators
shows the location of the on/off switch and indicators of the Quiet Cover MS control
panel.
Handling Temperature Alarms
In normal operation the Quiet Cover MS fan will turn off and on to maintain the temperature set
limits. The status light will be green when the fan is off. The status light will be orange when the
fan is on, but the temperature is still within set limits.
Both statuses are normal operating modes
.
If the air flow to the Quiet Cover MS is restricted or if the room temperature is above 35 °C (95 °F),
the red status indicator on the front panel lights and the temperature alarm sounds.
When this condition occurs, remove any items from around the Quiet Cover MS that could be
restricting air flow. Check that the fan is turning and that air flow to the Quiet Cover MS is not
restricted.
NOTE
It is important that the Quiet Cover MS not recirculate hot air. Always position the front end of
the Quiet Cover MS where it can draw cool air.
You cannot turn the alarm off. It will stop when the temperature returns to an acceptable level.
Figure 10.
Control panel
Power switch
Status light