Verifying the Installation
Verifying the Installation
Quiet Cover MS User Manual
19
Verifying the Installation
NOTE
When the G6011B Quiet Cover is plugged into an external power outlet, the Power Switch (front
lower left-hand corner of the Quiet Cover) will light up orange. This indicates that power is
connected to the Quiet Cover.
To verify that you installed the Quiet Cover MS correctly,
1
Position the power cords so they do not interfere with the kickstand (
).
2
Lock the casters to avoid unwanted movement of the Quiet Cover MS.
3
Turn the Power Switch to the off position.
4
Plug the G6011B Quiet Cover into an external power outlet. If the instrument is receiving
power, the Power Switch (
) should light up orange.
5
Turn the Power Switch on. If the instrument is turned on correctly, the Power Switch (
)
should light up green.
If the ambient temperature is < 35 °C, the temperature alarm and the
fan should be off
, and
the status indicator should light up green.
If the Quiet Cover internal temperature exceeds 35 °C, the fan should turn on, and the status
light will light up orange. This is the normal operating mode.
NOTE
In normal operating modes, the status light will be either green or orange, depending on if the fan
is on or not.