Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
4
Remove the Lid Top Cover
Remove the lid top cover.
Disconnect the Heater/Sensor Cable
1.
Trace the detector heater/sensor cable to the wiring harness near the
injection part cooling fan.
2.
Disconnect the cable at the connector.
Raise the Lid to the Service Position
1.
Open the lid. Locate the counterbalance cam in the left rear corner
under the lid. Loosen the screw on the right side of the cam. This allows
the stop plate to drop down.
Lid top cover, without
valve box accessory
Remove T-20 Torx™
screws (8 places)
Remove detector
cover, if desired
Remove vent tubing,
if present