Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
13
9.
Install the lid top cover.
Finishing Up
1.
Install the capillary adapter, if used.
2.
Restore the column connection.
3.
Restore carrier and other gases to the instrument.
4.
Restore power.
5.
Apply your normal operating pressures. Leak-check the manifold, back
panel, and column fittings.