Agilent G2615A Bar Code Reader

Agilent Technologies

User Information

Parts identification

4

Installation

5

About bar codes

7

Label and vial information

13

Applying labels

14

In case of difficulty

18

The Agilent G2615A Bar Code Reader (BCR) attaches to the tray 
of a 7683 Automatic Sampler. It reads labels attached to sample 
vials for positive sample identification.

The reader attaches directly to the 7683 tray. A provided 
spacer may be required to attach some trays.

The G2615A Bar Code Reader is not compatible with trays 
for 18596A, 18596B or 18596C samplers.

The 7683 tray must be delivered from Agilent with A.01.04 or 
greater firmware. Field upgrade to A.01.04 or greater is not 
adequate.

You can verify your tray firmware using the 6890N keypad and 
display. Select 

Options/Diagnostics/Instrument status

. Scroll 

to observe what firmware is installed on the G2614A tray. 
6890A and 6890 Plus GCs will not display the tray firmware. 
Contact your local Agilent office for assistance.

Summary of Contents for G2615A

Page 1: ...Agilent Technologies Agilent G2615A Bar Code Reader User Information...

Page 2: ...ad Wilmington Delaware 19808 1610 USA Safety Notices CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhe...

Page 3: ...rectly to the 7683 tray A provided spacer may be required to attach some trays The G2615A Bar Code Reader is not compatible with trays for 18596A 18596B or 18596C samplers The 7683 tray must be delive...

Page 4: ...1 identifies those parts in the installation kit that may not be easily recognized Figure 1 Parts identification Label applicator Vial plates Roll of labels Labeled vial Long screws Short screws 2 mm...

Page 5: ...style tray requires use of the spacer to mount the BCR 3 If the spacer is needed align the holes in the spacer with the mounting holes in the bottom of the tray Place the mounting bracket under the s...

Page 6: ...on 5 This completes bar code reader installation See the section on Automatic Samplers in your Agilent 6890 Series Gas Chromatograph user information to configure the bar code reader Figure 3 Connecti...

Page 7: ...BCR reads and decodes several different bar codes The most useful one for general laboratory use is the Standard 3 of 9 Code which encodes both letters and numbers and allows variable message length...

Page 8: ...characters dividing by 43 and using the remainder to select the checksum character If a checksum is used the character is placed after the message characters When a checksum is used the bar code reade...

Page 9: ...numbered positions are encoded with the five spaces associated with the preceding even character There is no character separator EAN JAN 13 codes The printed EAN JAN 13 symbol is divided into left and...

Page 10: ...l the left side pattern sets A and B If the first country of origin character is 0 the second country of origin character and the first five data characters will all be encoded using pattern set A AAA...

Page 11: ...the zero suppressed version and is used for small packages It requires that the number system character and at least four of the 10 data characters be 0 zero The Bar Code Reader can process both versi...

Page 12: ...e It can encode both letters and numbers plus a few punctuation marks and message length can be varied to suit both the amount of data to be encoded and the space available Interleaved 2 of 5 code is...

Page 13: ...meet the specifications in Figure 4 and Figure 5 respectively Figure 4 Label dimensions Figure 5 Vial dimensions 5 6 0 6 No label No label 20 5 1 0 or tape Maximum width measured at overlap 12 56 All...

Page 14: ...nt labels this is about 20 minutes Give the label bond a chance to anneal before you stress it whether mechanically or by placing it in a refrigerator or freezer The Agilent labels have a slight overl...

Page 15: ...ssume that Agilent labels are used 1 Remove the screws holding the side plate See Figure 6 2 Remove the side plate and discard the empty roller if any 3 Unwrap a new roll of labels Use care because th...

Page 16: ...8 Feed the end of the tape under the bottom plate and up through the large opening See Figure 7 9 Fold the tape and feed the end through the space between the bottom and top plates It will help to be...

Page 17: ...until the end of a label appears under the bottle 3 Hold the bottle in position let it rotate and continue to feed the label onto the bottle 4 Remove the bottle and press down the trailing edge of th...

Page 18: ...Labels or tape on the lower part of the vial may interfere with delivering the vial to its destination Labels must be applied smoothly with no bumps Ends must be securely fastened down Bar code label...

Page 19: ...o the vial may be carried at an angle and the base of the vial may miss its intended receptacle when lowered Labeled test vial A properly labeled test vial is provided with the bar code reader It is l...