Two vial plates are supplied with the applicator. The one
marked UNCAPPED VIALS is used when applying labels to vials
that have not been capped. Vials that have been capped will not
fit in this plate. A second plate, marked CAPPED VIALS, allows
for the extra length of the cap. While uncapped vials will fit in
this plate, they may not be properly positioned when the label is
applied.
Loading labels
The
following instructions assume that Agilent
labels are used.
1
Remove the screws holding the side plate. See
Figure 6
.
2
Remove the side plate and discard the empty roller (if any).
3
Unwrap a new roll of labels. Use care because the roll will
tend to unwind.
4
Place a new roll of labels on the three posts. The labels must
be on the outside of the roll and the roll must feed from the
bottom. See
Figure 6
.
5
Replace the side plate and the screws.
6
Verify that the correct vial plate (CAPPED or UNCAPPED) is
in place. If not, remove the four screws holding the top and
bottom plates, make the change, and replace the screws.
Figure 6
Loading labels
Do not let
tape uncoil
Side
Side plate
Vial plate