Inverted Magnetron Pirani Gauge FRG-700 FRG-702
12
Materials on the vacuum side
Vacuum connection
Measurement chamber
Feedthrough isolation
Internal seal
FRG-700
FRG-702
Anode
Ignition aid
Pirani measurement tube
Pirani filament
stainless steel
stainless steel
ceramic
FPM 75
Ag, Cu, soft solder (Sn, Ag)
Mo
stainless steel
Ni, Au
W
Internal volume
20 cm³
Pressure
10 bar (absolute),
limited to inert gases
Temperatures
Operation
FRG-700
FRG-702
Bakeout
Pirani filament
Storage
+5
… +55 °C
+5
… +150 °C
3)
150 °C (without electronics
and magnetic shielding)
120
C
-40
… +65 °C
Relative humidity
80% at temperatures
+31°C decreasing to 50%
at +40°C
Mounting orientation
any
Use
indoors only, altitude up to
2000 m
Degree of protection
IP 40