Chapter 1
49
Programming Fundamentals
Putting Multiple Commands on the Same Line
:TRIGger
[:SEQuence]
:EXTernal [1]
:SLOPe
POSitive
Bad Command
Good Command
PWR:ATT 40dB
POW:ATT 40dB
The short form of
POWER
is
POW
, not
PWR
.
FREQ:STAR 30MHz;MIX:RANG –20dBm
FREQ:STAR 30MHz;POW:MIX:RANG
–20dBm
The
:MIX:RANG
command is in the same
:SENSE
subsystem as
:FREQ
, but executing
the
:FREQ
command puts you back at the
:SENSE
level. You must specify
:POW
to
get to the
:MIX:RANG
command.
FREQ:STAR 30MHz;POW:MIX RANG
–20dBm
FREQ:STAR 30MHz;POW:MIX:RANG
–20dBm
:MIX
and
:RANG
require a colon to separate them.
:POW:ATT 40dB;TRIG:FREQ:STAR
2.3GHz
:POW:ATT 40dB;:FREQ:STAR
2.3GHz
:FREQ:STAR
is in the
:SENSE
subsystem, not the
:TRIGGER
subsystem.
:POW:ATT?:FREQ:STAR?
:POW:ATT?;:FREQ:STAR?
:POW
and
:FREQ
are within the same
:SENSE
subsystem, but they are two separate
commands, so they should be separated with a semicolon, not a colon.
:POW:ATT -5dB;:FREQ:STAR 10MHz
:POW:ATT 5dB;:FREQ:STAR 10MHz
Attenuation cannot be a negative value.