300
Chapter 5
Language Reference
[:SENSe]:SWEep Subsection
Sweep Time Mode
[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer
[:SENSe]:SWEep:TIME:AUTO:MODE?
Specifies the type of automatic coupling for the fastest sweep time at the current
settings.
Stimulus response
Spectrum analyzer
Factory Preset
and *RST:
SANalyzer
Front Panel
Access:
Sweep, Sweep Coupling SR SA
Time Gating Delay (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE:DELay <time>
[:SENSe]:SWEep:TIME:GATE:DELay?
Sets the delay time from when the gate trigger occurs to when the gate opens. This
is for
EDGE
triggering only.
Factory Preset
and *RST:
1
µ
s
Range:
0.3
µ
s to 429 seconds
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Delay
Time Gate Length (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE:LENGth <time>
[:SENSe]:SWEep:TIME:GATE:LENGth?
Specifies the gate time length in seconds; for
EDGE
triggering only.
Factory Preset
and *RST:
1
µ
s
Range:
0.3
µ
s to 429 seconds
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Length