Chapter 5
199
Language Reference
CALCulate Subsystem
CALCulate Subsystem
This subsystem is used to perform post-acquisition data processing. In effect, the
collection of new data triggers the CALCulate subsystem. In this instrument, the
primary functions in this subsystem are markers and limits.
NdBpoints
:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl>
:CALCulate:BWIDth|BANDwidth:NDB?
Selects the power level, below the peak of the signal, at which the signal
bandwidth will be measured by the markers.
:
CALCulate:BWIDth|BANDwidth[:STATe]
must be ON.
NOTE
To query the result of NdBpoints, use the command
:CALCulate:BWIDth|BANDwidth:RESult?
Factory Preset
and *RST:
–3 dB
Range:
–80 dB to –1 dB
Default Unit:
dB
Remarks:
Refer to :
CALCulate:BWIDth|BANDwidth[:STATe]
for an
explanation of this marker function.
Front Panel
Access:
Peak Search (or Search), N dB Points
NdBresults
:CALCulate:BWIDth|BANDwidth:RESult?
Returns the measured bandwidth at the power level defined by
:CALCulate:BWIDth:NDB
. –100 is returned if
:CALCulate:BWIDth|BANDwidth[:STATe]
is off, or when a result is not
available. Refer to
CALCulate:BWIDth|BANDwidth[:STATe]
for an
explanation of this marker function.
Range:
Real value less than the current frequency span
Default Unit:
Hz
Remarks:
When segmented sweep is on, a result will not be available
when the NDB marker crosses a segment boundary.
Front Panel
Access:
Peak Search (or Search), N dB Points