Chapter 2
21
Installation
Environmental Requirements
2.
Inst
allation
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy of the
analyzer, you must keep environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer or, for the rackmount type, by
forcefully air-cooling inside the rack housing. For more information on environmental
temperature to satisfy the specifications and measurement accuracy of the analyzer, see
“Specifications And Supplemental Information” in the
E5070B/E5071B User’s Guide
.
When the environmental temperature around the analyzer is kept within the temperature
range of the operating environment specification (See Section “Operating Environments”
on page 20), the analyzer conforms to the requirements of the safety standard.
Furthermore, under that temperature requirement, the analyzer still conforms to the
requirements of the safety standard even when placing the analyzer with cooling clearance
as follows:
Figure 2-2
Ventilation space at the installation
Requirements
Back
≥
180 mm
Sides
≥
60 mm (both right and left)