106 Register-Based Programming
Appendix B
For example, to connect both bank A and bank B to the analog bus, write a
“1” to bits 0 and 1 of the Tree Relay Control Register (base + 28
16
) to close
channel 90 (VSA) and channel 91 (VSB). All other bits must be set to “0”.
:ULWHD³´WRFORVHDFKDQQHOZULWHD³´WRRSHQDFKDQQHO
5HDGLQJWKHFKDQQHOELWLQGLFDWHVWKHVWDWHRIWKHUHOD\GULYHUFLUFXLWRQO\,WFDQQRWGHWHFWD
GHIHFWLYHUHOD\
base +
20
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Write
ch15 ch14 ch13 ch12 ch11 ch10
ch9
ch8
ch7
ch6
ch5
ch4
ch3
ch2
ch1
ch0
Read
base +
22
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Write
ch31 ch30 ch29 ch28 ch27 ch26 ch25 ch24 ch23 ch22 ch21 ch20 ch19 ch18 ch17 ch16
Read
base +
24
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Write
ch47 ch46 ch45 ch44 ch43 ch42 ch41 ch40 ch39 ch38 ch37 ch36 ch35 ch34 ch33 ch32
Read
base +
26
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Write
ch63 ch62 ch61 ch60 ch59 ch58 ch57 ch56 ch55 ch54 ch53 ch52 ch51 ch50 ch49 ch48
Read
base +
28
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Write
undefined
RTB
ch94
RTA
ch93
CS
ch92
VSB
ch91
VSA
ch90
Read
undefined
Tree Relay Control Bits
bit 0
VSA (ch90)
Connects the Voltage Sense H-L-G terminals of the Analog Bus to Bank A.
bit 1
VSB (ch91)
Connects the Voltage Sense H-L-G terminals of the Analog Bus to Bank B.
bit 2
CS (ch92)
Connects the Current Source H-L-G terminals of the Analog Bus to Bank B.
bit 3
RTA (ch93)
Connects the Reference Thermistor to Bank A for voltage sense.
bit 4
RTB (ch94)
Connects the Reference Thermistor to Bank B for current source.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com