ADM Flow Meter
Info mode
Agilent CrossLab CS ADM Flow Meter Operation Manual
17
07 Overflow
Warning
Measurement
exceeds instrument
range
Measure gas flow
within the acceptable
flow range
08 Cart Exp
Warning
Cartridge expired
Install a new cartridge
09 Low RTC Bat
Warning
RTC (real time
clock) backup
battery low
Contact Agilent
Technical Support
10 Op Temp
Error
Out of operating
temperature range
condition
(0 to 45 °C)
Operate the flow meter
within the acceptable
temperature range
11 Op Pres
Error
Out of operating
atmospheric
pressure range
Operate the flow meter
within the acceptable
atmospheric pressure
range
12 Ver
Mismatch
Error
Firmware version
mismatch with
cartridge version
Upgrade firmware to
latest version
13 Invalid Cart
Error
Unable to identify
cartridge types
Contact Agilent
Technical Support
Table 2
Error and warning code descriptions (continued)
Error/
Warning
code
Type
Description
Troubleshooting