Technical Information
Typical Layout Diagram
17
Turbo-V 81-T User Manual / 87-900-984-01 (G
)
225/234
Connection B - Fore-Vacuum Pump
A flange KF 16 NW is available to connect the Turbo-V 81-T pump to
the fore-vacuum pump. A hose or vacuum approved pipe can be used.
If a rigid pipe is used, any vibration generated by the mechanical
pump must be eliminated through the use of bellows.
NOTE
The Turbo-V 81-T pump is characterized by its high compression ratio also for oil
vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.
Connection C - Electrical
Figure 34
The turbopump is connected to the controller through an 6-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pin E
is connected to the pump ground, pins A and F reserved for Agilent.
If the temperature sensor is disconnected, the pump will not start. To
prevent damage to the pump when the temperature exceeds 60 °C,
the sensor automatically cuts out the power supply.
Summary of Contents for 9698905
Page 3: ...Turbo V 81 T Turbo V 81 T User Manual 87 900 984 01 G 3 234 Turbo V 81 T...
Page 4: ...Turbo V 81 T 4 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 96: ...7 Istruktionsbog Bortskaffelse 96 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 130: ...10 Ohjek sikirja H vitt minen 130 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 142: ...11 Felhaszn l i K zik nyv Megsemmis t s 142 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 154: ...12 Podrecznik Instrukcji Przetworstwo odpadow 154 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 176: ...14 N vod na Obsluhu Likvid cia 176 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 188: ...15 Priro nik za Navodila Odlaganje opadkov 188 234 Turbo V 81 T User Manual 87 900 984 01 G...
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......