17 Technical Information
Connection C - Electrical
284/294
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
Connection C - Electrical
Figure 42
The turbopump is connected to the controller through an 8-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pin E
is connected to the pump ground, pins G and H are not connected,
pins A and F reserved for Agilent.
If the temperature sensor is disconnected, the pump will not start. To
prevent damage to the pump when the temperature exceeds 60 °C,
the sensor automatically cuts out the power supply.