Measurement
System
Conguration
With
an
8757
series
scalar
network
analyzer
using
an
85025A/B/D/E
detector
,
system
conguration
requires
special
attention.
A
C
mode
is
the
default
state
of
the
analyzer
system
and
there
are
no
further
requirements
to
initiate
a
measurement.
However
,
to
enable
DC
mode
operation,
a
series
of
keystrokes
is
required.
DC
Detection
mode
DC
detection
oers
greater
power
measurement
accuracy
and
the
ability
to
characterize
oscillators
and
modulation-sensitive
devices
.
Figure
3-2
depicts
a
typical
measurement
setup
for
0.01to
50
GHz,
using
an
Agilent
8350B
sweep
oscillator/RF
plug-in
as
the
source
.
1
.
On
the
analyzer
,
press
4
PRESET
5 .
Connect
the
detector(s).
2.
DC
detection
mode
must
be
selected.
On
the
analyzer
,
press
4
SYSTEM
5
and
select
NNNNNNNNNNNNNNNNNNNNNNN
MODE
DC
.
When
the
NNNNNNNNNNNNNNNNNNNNNNN
MODE
DC
softkey
is
selected,
the
source's
square
wave
modulation
is
automatically
switched
OFF
.
A
ccurate
DC
Measurements
Zeroing
the
Detector
When
you
make
DC
detection
measurements
,
it
is
important
to
perform
this
detector
zeroing
procedure
to
compensate
for
the
eects
of
DC
drift
and
temperature
uctuations
.
This
zeroing
procedure
eliminates
small
DC
voltages
from
the
diode
detector
that
would
otherwise
cause
amplitude
measurement
errors
at
low
(40
dBm)
power
levels
.
Zeroing
also
establishes
the
displayed
noise
level
with
no
RF
signal
applied
(the
system's
noise
oor).
A
utozero
Pressing
the
autozero
softkey
NNNNNNNNNNNNNNNNNNNNNNN
AUTOZRO
switches
OFF
the
source
RF
signal
output
and
automatically
zeros
the
detector
.
Agilent
85025A/B/D/E
Operation
3-3
Summary of Contents for 85025A
Page 10: ......
Page 12: ...Figure 1 1 85025A B D E Detector 1 2 General Information Agilent 85025A B D E...
Page 24: ......
Page 37: ......
Page 66: ......
Page 68: ......
Page 71: ...Figure 6 1 Module Exchange Program Agilent 85025A B D E Replaceable Parts 6 3...
Page 72: ......
Page 78: ......
Page 82: ......
Page 96: ......