Measurement Accuracy
Measurement Accuracy
To ensure that you obtain the specified accuracy, you must perform a plug-in
module vertical calibration. The calibration must also be performed when you
move a plug-in module from one slot to another, or from one mainframe to
another. Refer to Chapter 3, “Calibration Overview” for information on per-
forming a plug-in module vertical calibration.
C A U T I O N
The Agilent 83486A input circuitry can be damaged when the
total
input power
levels 10 dBm (10 mW) on the optical channel or ±2 V + peak ac
(+16 dBm) on the electrical channel. To prevent input damage, this specified
level must not be exceeded.
Measurement accuracy—it’s up to you!
Fiber-optic connectors are easily damaged when connected to dirty or damaged cables
and accessories. The Agilent 83486A’s front-panel INPUT connector is no exception.
When you use improper cleaning and handling techniques, you risk expensive instru-
ment repairs, damaged cables, and compromised measurements.
Before you connect any fiber-optic cable to the Agilent 83486A, refer to “Cleaning Con-
nections for Accurate Measurements” on page 5-9.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 83486A
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 36: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 89: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...