3-11
Calibration Overview
User Calibrations—Optical and Electrical
Plug-in Module Vertical Calibration
The plug-in module vertical calibration is for both optical and electrical mea-
surements. It allows the instrument to establish the calibration factors for a
specific plug-in when the plug-in is installed in the mainframe. The plug-in cal-
ibration factors are valid only for the specific mainframe slot in which it was
calibrated. The plug-in vertical calibration establishes vertical accuracy.
A plug-in vertical calibration should be done if:
• The mainframe has cycled power.
• The plug-in has been repaired, reprogrammed, or removed from the
mainframe.
• The instrument’s operating temperature has changed and remains more than
5
°
C from the temperature at which the Plug-in calibration was performed.
To obtain the best measurement results, it is recommended that a user verti-
cal calibration be performed after every 10 hours of continuous use or if the
temperature has changed by greater than 2
°
C from the previous vertical cali-
bration.
To view the temperature change
This procedure displays the temperature change that the instrument has
undergone since the last Plug-in Vertical Calibration.
1
Press the front-panel channel
SETUP
key.
2
Press
Calibrate
and then
Cal status on
.
The current plug-in
∆
Temp value is listed for each installed module.
To perform a plug-in module vertical calibration
1
Remove any front-panel connections from electrical channels.
2
Cover the optical inputs for the optical channels.
3
Press
Utility
,
Calibrate
..., and then
Calibrate plug-in
....
4
Select the plug-in module to be calibrated, press
1 and 2
or
3 and 4
.
5
Press
Start cal
to start the calibration.
6
Follow the on-screen instructions.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 83486A
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 36: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 89: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...