Procedure: Testing Resistance of the Heated Zone Heaters and PRT’s 

WARNING: Before proceeding, turn off the main power switch and unplug 
the power cord. 

Caution: Make sure you are properly grounded with an ESD strap before 
continuing. 

 

 

Remove the GC panels required for access to the zone heater/sensor connector to be tested. 

 

Remove the connector for the heated zone sensor to be tested

 – use an ohm meter to 

measure the resistance of the PRT in question.  

 

For heated zones the PRT is connected to the white wires on pins 2 and 3 of the connector. 

 

The resistance of the PRT should correspond to those in the “

Table of PRT Values

” 

 

The heater resistance can also be measured between the red wires on pins 1 and 4 – typically 
23 Ω for a 70 watt heater (see table following). 

 

With the 

connector removed from the GC

, there should also be no connection to ground on 

any of the four pins – this would indicate a possible shorted sensor or heater. 

 

                                             

 

Table: PRT Resistance Values: 

 

 

Pin 4 

– Heater Ground 

Pin 3 

– Sensor Ground 

Pin 2 

– Temp Sense 

Pin 1 

– Heater Control 

Summary of Contents for 7890 Series

Page 1: ...C module installed in the same position If you want to use a Detector Connection you cannot have an associated EPC module or Detector Logic board installed 7890 heaters must be between 10 and 180 watts There is no configurable total power limit If total power F or B channel 220 watts power priority to zones is 1 Inlet 2 Aux 3 Det Thermal Aux Zone Configuration Menus NOTES Before turning on the GC ...

Page 2: ...ser Configurable Heater Move the cursor to User Configurable HTR select Again you will see Again Select From here you can input custom configuration values for your User Configuraeable Heater When all settings have been entered you can finally Power Cycle the GC to make your changes take effect Configuring a Flame Photometric Heater Configuration Access Menu Description Front or Back Detctor The F...

Page 3: ... measure the resistance of the PRT in question For heated zones the PRT is connected to the white wires on pins 2 and 3 of the connector The resistance of the PRT should correspond to those in the Table of PRT Values The heater resistance can also be measured between the red wires on pins 1 and 4 typically 23 Ω for a 70 watt heater see table following With the connector removed from the GC there s...

Page 4: ...ater Type K Thermocouple Resistance cold 13 9 PTV 6890 7890 G2617 60506 Whole Inlet Assembly 150 Watt Heater Type K Thermocouple Resistance cold 13 9 VI G2319 60503 60 Watt Heater 100 PRT 26 27 FID G1530 67950 70 Watt Heater 100 PRT 23 TCD G1530 67950 70 Watt Heater 100 PRT 23 NPD G1530 67950 70 Watt Heater 100 PRT 23 uECD G1533 60625 70 Watt Heater 100 PRT 23 FPD G1535 60610 Block G1535 60620 Xfe...

Reviews: