Maintenance and Troubleshooting
48
Agilent 700 Series ICP Optical Emission Spectrometers User’s Guide
Axial Induction Coil
Figure 10.
Induction coil — axial where 1. Torch, 2. Induction coil and 3. Torch
clamp
Pitch of Axial Induction Coil
Figure 11.
Pitch measurement — an axial induction coil must have a pitch (
↔
)
of 7.5 ±0.5 mm
NOTE
Pitch is measured from the center of one coil to the next, as shown above by the
arrows between the dotted gray lines, and
not
the gap between adjacent coils.