Introduction
Agilent 700 Series ICP Optical Emission Spectrometers User’s Guide
19
Environmental Conditions
The instrument is suitable for indoor use
only
and is classified
suitable under the following categories (EN 61010-1):
Installation category II
Pollution degree 2
Equipment class I
Temperature Control
For
optimum analytical performance
, it is recommended that the
ambient temperature of the laboratory be between 20 and 25 °C (68
and 77 °F).and be held constant to within
±
2 °C (
±
3.6 °F) throughout
the entire working day.
Electrical Power Supplies
For electrical requirements, refer to your Agilent 700 Series ICP
Optical Emission Spectrometers Site Preparation Guide.
Fuses
1FS1 and 1FS2
T6.3 AH 250 V, IEC 127 sheet 5, 5 x 20 mm
1CB1
Circuit breaker 30 A Fast trip
NOTE
For safety reasons, any other internal fuse or circuit breaker is not operator
accessible, and should only be replaced by Agilent-authorized personnel.
Fuse information on the rear of the instrument is the most up to
date.