Installing Sampling Valves on an Agilent 6850 Gas Chromatograph
Installing Sampling Valves
9
Installing a Sampling Valve
The sampling valve is installed in a heated valve box on top
of the GC. See
. The box and valve is a
factory- assembled and leak- tested unit. Installation does not
require any disassembly of the unit. See the
6850 Gas
Chromatograph User Information
CD- ROM.
The valve is plumbed through an existing inlet, either
capillary or packed.
Table 3
Installing a sampling valve
Step
Action
Notes
1
Remove the lid top cover.
a
Remove vent tubing, if present.
b
Remove detector cover.
c
Remove 8 T-20 Torx screws.
d
Remove top cover.
2
If the valve is to replace a
capillary
inlet
, remove the insert weldment
and split vent trap.
If the valve is to replace a
packed
inlet, continue at Step 3.
a
Use the inlet wrench to release the
large nut on the inlet insert weldment.
b
Lift the inlet insert weldment clear of
the inlet.
Detector cover
T-20 Torx screw
(1 of 8)
Vent tubing,
if present