18
Installing Sampling Valves
Installing Sampling Valves on an Agilent 6850 Gas Chromatograph
2
Connect the control cable.
•
Connect the cable from the control
valve to the wiring harness.
3
Install the modified top cover.
a
Place the modified lid top cover
supplied in this kit on top of the lid.
b
Secure with eight screws.
4
Connect the external actuator air
supply tubing.
a
Connect the air source to the fitting on
the back of the lid (the control valve).
b
Set the source delivery pressure at 310
to 345 kPa (45 to 50 psi).
•
Valves are gas-driven, usually by
compressed air. There is no high
purity requirement for this gas. The
supply pressure may surge as the
valve operates, so the actuator gas
must not come from the FID air.
5
Finishing up.
a
Restore carrier and other gases to the
instrument.
b
Restore power.
c
Apply your normal operating
procedures. Leak-check all new
connections.
Table 5
Installing an air control valve for a rotary-type sampling valve (continued)
Step
Action
Notes
Harness
Connector