User Connections
84
A
B
From 1 to 16 direct supplies may be connected to 1 controller GPIB interface.
Tighten connector thumbscrews by hand. Do not use a screwdriver.
Do not stack more than 3 connectors on a GPIB receptacle.
GPIB cable (see
Accessories
in Chapter 1)
From 1 to 15 linked supplies may be connected to 1 direct supply.
Either receptacle (Jl or J2) may be used as an input or an output.
Serial Link Cable (see
Accessories
in Chapter 1), 2 meters. 1 is supplied.
Maximum total length of all GPIB cables (including controller) not to exceed 20 meters.
Use caution with individual lengths over 4 meters.
Maximum total length of all serial cables not to exceed 30 meters.
1.
2.
3
.
NOTES:
A direct power supply is connected to the controller interface and must have a unique primary GPIB bus address.
The stand-alone configuration uses only direct supplies connected to the controller interface.
The linked configuration uses 1 or more linked power supplies connected to each direct supply. Each linked supply has
a unique secondary GPIB bus address and derives its primary address from the direct supply.
Figure 4-6. Controller Connections