General Information 33
Table 1-4b. Supplemental Characteristics for Series 668xA (continued)
1
Parameter
Agilent Model Number
6680A
6681A
6682A
6683A
6684A
Remote Sensing Capability
Voltage Drop Per Lead:
Load Voltage:
Up to 1/2 of rated output voltage.
Subtract voltage drop in load leads from specified output
voltage rating.
Load Regulation:
Degradation due to load lead drop in--output:
∆
mV (regulation) = Vdrop(R
sense
-)
Degradation due to load lead drop in + output:
∆
mV (regulation) = V
drop
(R
sense
+) + 2V
drop
(V
rating
)/(V
rating
+ 10 V)
where R
sense
_ and R
sense
+
are resistances of respective sense leads.
Maximum Reverse Voltage Current Sink Capability: *
With ac input power applied and the dc output reverse biased
by an external dc source, the supply will continuously
withstand without damage a current equal to its output
current rating.
* Current must be limited by user's external dc source.
Load Voltage:
Subtract voltage drop in load leads from specified output
voltage rating.
Maximum Input Power:
7350 VA,
6000 W,
160 W (with no load)
Maximum AC Line Current Ratings
Range 1
Rms line current:
21.4 A (27.7 A) ***
Line fuse:
30 AM
Range 2
Rms line current:
10.7 A (14.4 A) ***
Line fuse:
16 AM
*** Includes 5% unbalanced voltage phase condition.
Output Voltage Programming Response Time**
Programming Rise/Fall Time
(time for output to change from 90 % to 10% or from 10% to 90% of its total
excursion):
***
9 ms
12 ms
45 ms
60 ms
60 ms
Full-load programming speed up/down time
(time for output to settle within 4 LSBs of the final value):***
27 ms
35 ms
140 ms
185 ms
185 ms
No-load downprogrammiug discharge time
(time for output to fall to 0.5 V when programmed from full voltage to
zero volts):
90 ms
100 ms
475 ms
650 ms
575 ms
**
All values exclude command processing time.
***
With full resistive load = V
RATED
/I
RATED
Notes
:
l
For Performance Specifications, see Table l-4a.