38 Installation
Turn-On/Selftest
Turn on the Electronic Load using the LINE switch on the front panel and observe the display. Immediately after turn-on,
the Electronic Load undergoes a selftest that checks the GPIB interface board as well as the input circuitry of the installed
modules. All of the front panel LCD segments are momentarily activated. When selftest completes, the display should
appear about the same as the one shown in Figure 3-6 with the CC annunciator being on.
Figure 3-6. Front Panel Display
After the Electronic Load has passed selftest, connect a power supply to the Electronic Load to test the input circuits as
described under "Power Test".
If the Electronic Load fails any portion of the selftest, one of the following error numbers may briefly appear on the display:
GPIB Errors
Channel Errors
Display
Description
Display*
Description
ERROR 1
RAM failure
1 ERROR 100
Self test error
ERROR 2
ROM failure
1 ERROR 101
Secondary RAM failure
ERROR 3
GPIB failure
1 ERROR 102
Secondary ROM failure
ERROR 4
Internal trigger failed
1 ERROR 103
Secondary timer trigger failed
ERROR 5
Line trigger failed
1 ERROR 104**
Calibration EEprom failed
ERROR 6
Line trigger failed
1 ERROR 105
Main DAC high
ERROR 7
EEprom failed
1 ERROR 106
Main DAC low
1 ERROR 107
Transient DAC high
1 ERROR 108
Transient DAC low
* The applicable channel number is displayed for Multiple Electronic Loads starting with the lowest numbered channel.
Error numbers are not related to the negative numbers returned by the SYST:ERR? query.
** Requires recalibration
Another indication that a selftest failure has occurred is if the ERR annunciator on the display remains on after selftest
completes.
If the Electronic Load has failed selftest, the mainframe or module to the nearest Agilent Sales and Service Office for repair.
Summary of Contents for 6050A
Page 10: ......
Page 14: ......
Page 32: ......
Page 46: ...46 Installation Figure 3 12 Local Sensing Figure 3 13 Remote Sensing...
Page 47: ...Installation 47 Figure 3 14 Parallel Operation Figure 3 15 Zero Volt Loading...
Page 48: ......
Page 55: ...Local Operation 55 Figure 4 2 Recommended Programming Sequence...
Page 64: ......
Page 69: ...Remote Operation 69 Figure 5 1 Remote Programming Flowchart Sheet 1...
Page 70: ...70 Remote Operation Figure 5 1 Remote Programming Flowchart Sheet 2...
Page 74: ......
Page 78: ...78 Calibration Figure 6 2 Calibration Flowchart for A Modules...
Page 79: ...Calibration 79 Figure 6 2 Calibration Flowchart for A Modules continued...
Page 80: ...80 Calibration Figure 6 2 Calibration Flowchart for A Modules continued...
Page 85: ...Calibration 85 Figure 6 3 Calibration Flowchart for B Modules...
Page 86: ...86 Calibration Figure 6 3 Calibration Flowchart for B Modules continued...
Page 87: ...Calibration 87 Figure 6 3 Calibration Flowchart for B Modules continued...