CI Maintenance
To Disassemble the CI Source
238
5977B Series MSD Operating Manual
To Disassemble the CI Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
•
Nut driver, 5.5 mm (8710-1220)
•
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view
and CI source parts list
page 239 while using this procedure.
1
Remove the CI source. (See
2
Remove the filaments. (See
“To Remove the CI Source Filament”
3
Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts.
4
Disassemble the repeller assembly by removing the ceramic insulator from
the repeller.
5
Remove the setscrew securing the lenses to the source body.
6
Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens.