8
CI Maintenance
To Remove the CI Source
5977B Series MSD Operating Manual
235
To Remove the CI Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Tweezers (8710-2460)
Procedure
The video shows wiring connections for the non-HES model MSD.
1
Vent the MSD. (See
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
page 176.)
3
Use tweezers to disconnect the wires from the CI source. Do not bend the
wires any more than necessary. (See
“To Connect/Disconnect non-HES
Model Wiring from the CI Source”
EI HES Model Wiring to the CI Source”
Trace the wires for the ion source heater and temperature sensor to the ceramic
source board, and disconnect them there. (See
Model Wiring to the CI Source”
4
Remove the two large thumbscrews that hold the ion source in place.
5
Pull the ion source out of the source radiator, and place it in its storage container.
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Make sure you use an antistatic wrist strap, and take other antistatic
precautions before touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......