8
CI Maintenance
To Remove the CI Source Filament
5977B Series MSD Operating Manual
247
To Remove the CI Source Filament
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Tweezers (8710-2460)
Procedure
1
Vent the MSD. (See
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
page 176.)
3
Remove the CI source. (See
4
Remove the screw holding the filament to the CI source body. (See
.)
5
Slide the filament off the CI source assembly. (See
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.