To Remove the EI HES filaments
192
5977B Series MSD Operating Manual
To Remove the EI HES filaments
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Cloths, clean (05980-60051)
•
T6 Torx driver, (8710-2548)
Procedure
1
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Remove the ion source and place it on a clean cloth on your work surface.
(See
4
Remove the screw securing the filament block to the source mount. (See
.)
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
Figure 62 Replacing the dual filament
Source mount
Screw
Filament block
Dual filament