7
General Maintenance
To Remove the EI HES
178
5977B Series MSD Operating Manual
To Remove the EI HES
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Tweezers (8710-2460)
Procedure
1
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Remove the two large thumbscrews that hold the ion source in place. (See
4
Disconnect the wires from the EI HES. (See
bend the wires any more than necessary. (See
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.