Chapter 3: Testing Performance
Analog Bandwidth - Maximum Frequency Check
3–17
Connections
Procedure
1
Preset the power meter.
2
Ensure that the power sensor is disconnected from any source and zero the meter.
3
Connect the power sensor to the power meter’s Power Ref connector and calibrate the
meter.
4
Make the connections to scope channel 1 as shown in the connection diagram above.
5
Set up the Power Meter to display measurements in units of Watts.
6
Press Default Setup, then configure the scope as follows:
a
Ensure Channel 1 is displayed and all other channels are turned off.
b
Set the vertical sensitivity of channel 1 to 10mV/div.
c
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
Notes
• Connect output 1 of the 11667B splitter to the scope Channel n input directly using the 54855-67604
adapter, without any additional cabling or adapters.
• Connect the power sensor directly to output 2 of the power splitter without any additional cabling or
adapters.
• Minimize the use of other adapters.
• Ensure that SMA and 3.5 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 3.5 mm
5 in-lbs (56 N-cm) for SMA
Power Meter
5485x Scope Under Test
Microwave
Generator
RF
Out
Ch
an 1
Chan 2
Chan 3
Chan 4
E
4
41
3A
11667B
54855-67604
Click here and
enter 16E-9
Summary of Contents for 54853A
Page 4: ...iv Rear panel with option 017 Removable hard drive ...
Page 6: ...vi ...
Page 10: ...Contents 4 ...
Page 24: ...1 14 ...
Page 44: ...2 20 ...
Page 172: ...6 38 ...
Page 177: ...Chapter 7 Replaceable Parts Exploded Views 7 5 Exploded Views Front Frame and Front Panel ...
Page 178: ...Chapter 7 Replaceable Parts Exploded Views 7 6 Fan and Acquisition Assembly ...
Page 181: ...Chapter 7 Replaceable Parts Exploded Views 7 9 Sleeve and Accessory Pouch ...
Page 187: ...8 Block Level Theory 8 3 Acquisition Theory 8 5 Theory of Operation ...
Page 192: ...8 6 ...