Chapter 4: Calibrating and Adjusting
To check the power supply
4–5
4
Connect the common lead of the voltmeter to the GND test point.
5
Connect the positive lead of the voltmeter to the +5.1 V test point.
6
Verify that the +5.1 supply voltage is within limits as shown in the following table:
Table 4-1
Power Supply Voltage Limits
7
Repeat steps 4 through 6 for each of the other supply voltages.
If any supply voltage is not within specifications, see chapter 5, “Troubleshooting.”
Supply Voltage Specification
Limits
+5.1 V
±
0.1 V
+5.0 V to +5.2 V
-5.2 V
±
0.1 V
-5.1 V to -5.3 V
+12.2 V
±
0.3 V
+11.9 V to +12.5 V
-12.2 V
±
0.3 V
-11.9 V to -12.5 V
+15 V bias
+14 V to +16 V