Chapter 6: Replacing Assemblies
To reset the attenuator contact counter
6–30
To reset the attenuator contact counter
1
Turn off the Infiniium oscilloscope.
2
Connect a mouse and a keyboard to the oscilloscope.
3
Insert the service disk in the floppy drive and turn on the oscilloscope.
For units equipped with a 120 Mbyte floppy drive, do not insert the service disk until
the message:
Starting Windows 9X...
appears on the screen.
4
The oscilloscope will boot into the Windows® operating system.
5
Press Control-Esc on the keyboard to access the Windows® Start menu.
6
Open the Programs group and click the Infiniium Scope icon.
7
When the oscilloscope application comes up, click the mouse icon in the upper right
corner of the screen.
8
Click the Utility pull down menu, then click Self Test. In the Self Test screen click
Select Service Extensions.
9
Click the Self Test pull down menu, scroll down and select Atten Set.
Figure 6-28
Self Test